Cleaning and maintenance
14
4 Cleaning and maintenance
Your MX2/UMX5 automated balance does not require laborious maintenance. For correct operation and to ensure correct results simply
carry out the cleaning and maintenance tasks described below. For information about general cleaning tasks for the balance, please
refer to Chapter 13 of the operating instructions for AX and MX/UMX balances.
Clean the components of the weighing chamber as follows:
Switch off the balance.
Tilt the locking rocker (A) of the weighing chamber cover to the right and remove the
weighing chamber cover by pulling it forwards.
First carefully remove the weighing sample holder (B) and then the weighing pan (C) by
lifting it out of its guide. We recommend using the special tool supplied with the balance
for this (special tweezers).
Then remove the drip tray (D) by lifting it out of the weighing chamber.
Clean the items (they are made from stainless steel X2CrNiMo17-12-2/1.4404).
In particular, handle the weighing sample holder and the weighing pan with the greatest
care and ensure that the fine holding prongs for the weighing sample are not bent.
We recommend cleaning all items in an ultrasonic bath.
Replace the items in the weighing chamber in reverse order. When inserting the weighing
pan, make sure that it is in the correct position: the fine holding prongs for the weighing
sample must be towards the back, positioned towards the weighing sample holder.
In addition, when inserting the weighing pan and the weighing sample holder make sure
that the items are positioned correctly in the slot in the relevant guide. Do not forget to
lock the weighing chamber cover (tilt rocking locker to the left).